Pressure and level transmitter L3 ANDERSON-NEGELE
Pressure and level transmitter L3 ANDERSON-NEGELE

We are not the manufacturer, authorised distributor or authorised representative of the ANDERSON-NEGELE brand in South Africa. The private labels and part numbers shown on this website are the property of their respective owners.

ANDERSON-NEGELE L3

Manufacturer

ANDERSON-NEGELE

Country

Germany

Name

L3

Category

Level Sensors

Type

Pressure and level transmitter

Stock availability:

On request

Price: On request

Order

Several positions are needed?

Send your request on email:

Worldwide delivery services:
fedex dhl ups
Ways of payment:
wire transfer paypal

Specifications L3

The Anderson-Negele L3 Pressure and Level Transmitter has been designed to measure process pressure or hydrostatic level in hygienicprocess applications. The state-of-the-art temperature compensation reduces errors associated with process temperature changes with improved zero stability reduces sensor interaction. The graphical user interface makes set-up and programming easy by directly aligning to the HARTTM DD menu structure. The field repairable and reconfigurable design allows the user to change the display orientation, add a remote cable, or replace a component in the field without impact to accuracy. Key benefits:
  • Improved temperature compensation
  • High accuracy also with turndown up to 1:10
  • Accurate level measurement also in small containers and with rapidly changing process temperatures
  • Modular design for individual configuration, with Smart Replace Design (easy component exchange in the field)
  • Integrated tank linearisation with density compensation for high-precision output of volume or mass (kg, litres, percent…)
  • Easy adjustment of zero and span

Features L3

Process ConnectionCLEANadapt | Milchrohr | DRD | SMS Liner | Tri-Clamp | CPM | Endress & Hauser Universal
Insertion lengthStainless steel 1.4404 (316L) Ra≤0,8 μm | Diaphragm 1.4404 (316L), Ra≤0,64 μm
Measuring range-1…35 bar Relative
Accuracy≤ 0.10% of the calibrated range
Available versionsCompact | Remote
Temperature range process-18…110 °C
Temperature range CIP/SIP135 °C / max. 60 min.
Communication4…20 mA | HART 7.0
The values shown are the performance limits within the product line. The product-specific data for each individual configuration can be obtained from the product information